Focusing X-rays Using Micropore Optics (MPO)
Description
This paper investigates the effects of Micropore Optics (MPOs) X-ray detected by MALTA. In this research, Photonis micro-Xray source was used. This X-ray source was set to have maximum X-ray photon energy of 12 KeV. Experiments have been done to determine the effects of source-detector distance and MPO-source distance to the X-ray beam. The MPOs affected the hits detected by MALTA and also the divergence of the photons. A single MPO can decrease the number of hits by a factor of 10$^{1}$. The divergence of the photons are analyzed by computing the standard deviation and divergence angle. Furthermore, three set-ups were also evaluated to determine which is the most effective in focusing X-ray photons.
Files
Focusing Micro-Xrays.pdf
Files
(4.0 MB)
| Name | Size | Download all |
|---|---|---|
|
md5:0839f5a1a8044531b71c9f61fe1224cd
|
4.0 MB | Preview Download |
Additional details
Identifiers
- CDS Report Number
- CERN-STUDENTS-Note-2022-012
CERN
- Department
- EP - Experimental Physics Department
- Experiment
- ATLAS