Published May 15, 2020
| Version v1
Thesis
Open
Characterisation of the new tracker sensors for the CMS experiment at CERN
Description
I-V and C-V curves are crucial measurements required to characterise Silicon sensors. They must be performed at reception and at several steps of particle detector module assembly to spot potential damages. High-voltage (1kV) biasing of those sensors and accurate current measurements (50??A max) are mandatory for sensor characterization. High-voltage electrometers and LCR meters with proper decoupling are the typical instruments used to perform those measurements. Those instruments are rather expensive (10k euros range)and are often over-specified for this precise task. We developed a system tailored to perform both I-V and C-V measurements of sensors in one compact, low-cost instrument.The instrument is built around a PCB that implements a low ripple (100mV) 1kV high voltage power supply, a high-side AC/DC current measurement circuit and a capacitance measurement circuit. The instrument is coupled to an external signal generator, a computer that extracts electrical parameters and the I-V/C-V curves, and a Programmable System-On-Chip (PSOC) which handles source control in addition to current, voltage and voltage/current phase measurements. On top of that, temperature and relative humidity are monitored. Finally, low-voltage DC regulators can power readout electronics of fully assembled modules to provide a standalone test station usable in sensor/module characterisation or in thermal cycling.
Files
TS2020_010_2.pdf
Files
(14.7 MB)
| Name | Size | Download all |
|---|---|---|
|
md5:9c7d23b0a91e938870dc1bc3fdf72f89
|
14.7 MB | Preview Download |
Additional details
Identifiers
- CDS
- 2791112
- CDS Report Number
- CMS-TS-2020-010
- CDS Report Number
- CERN-THESIS-2020-385
CERN
- Department
- PH
- Programme
- No program participation
- Accelerator
- CERN LHC
- Experiment
- CMS