In the Picosec project micro pattern gaseous detectors with a time resolution of some ten picoseconds are developed. The detectors are based on Micromegas detectors. With a cherenkov window and a photocathode the time jitter from different position of the primary ionization clusters can be substituted. This reports describes the beam setup and measurements of different Picosec prototypes. A time resolution of under 30 ps has been measured during the test beam. This report gives an overview of my work as a Summer Student. I set up and operated a triple-GEM tracker and a trigger system for the beam. During the beam I measured different prototypes of Picosec detectors and analysed the data.